Correspondence
Unsuitable blood glucose measuring devices in neonatal or paediatric acute care
Dankwart Friedrich Wittenberg, Alta Terblanche, Izelle Smuts, Johannes Opperman
Abstract
A newborn baby with persistent neonatal jaundice developed severe septicaemia. Blood glucose monitoring using Accuchek(R) revealed persistently high values, but laboratory blood glucose testing showed no instance of true hyperglycaemia. She was found to have galactosaemia.
The Accu-chek® glucose meter does not discriminate between blood galactose and glucose.
Authors' affiliations
Dankwart Friedrich Wittenberg, University of Pretoria
Alta Terblanche, Pretoria Academic Hospital
Izelle Smuts, University of Pretoria
Johannes Opperman, University of Pretoria
Keywords
Galactosaemia; Test Strip Blood Glucose; Glucose Dehydrogenase
Cite this article
South African Medical Journal 2007;97(12):1231.
Article History
Date submitted: 2007-05-31
Date published: 2007-12-14
Article Views
Abstract views: 3341
Full text views: 1334
Comments on this article